Application Note
SpectraSensors SS2000 Single Channel Analyzer for CO2 is extremely reliable and tailored for the needs of the natural gas industry. The sensor measures gas using a patented Tunable Diode Laser Diode (TDL) to determine the concentration of the gas without coming into physical contact with the stream.
RAPID RESPONSE TIME The SS2000 analyzer takes four measurements per second with a laser and detector and immediately averages the results. Because there is no contact with the gas, real-time measurements are not hampered by wet-up or dry-down times as with surfaced-based sensors.
RELIABLE Trustworthy measurements are vital to natural gas pipeline and processing companies. Independent studies have proven that the SS2000 results are highly correlated with those of chilled mirrors. However, chilled mirrors require skilled experts to operate and the results are highly scattered (large standard deviation).
Uncertain measurements can be extremely costly. Additional processing of dehydration costs, upset conditions, shut-ins and inconsistent process results may be caused by sensors that do not perform properly. The SS2000 is the first to offer truly reliable measurement and simple operation.
LONG LIFE The TDL sensor does not come into contact with the sample gas stream. The result is a sensor which does not suffer from contamination or drift due to vapor impurities such as glycol, methanol or amines.
LOW COST OF OWNERSHIP Operating costs are dramatically reduced by eliminating the cost of consumables, extra sensor heads, labor and overhead associated with excessive maintenance.
The SS2000 dramatically reduces intangible but real costs associated with unreliable gas measurements. By eliminating added processing steps, detecting poor gas quality and the possibility of costly damage to equipment that can result from sensors that produce incorrect data.
Data Sheet
Performance |
|
|
Target Components |
CO2 in Natural Gas |
|
Typical Measurement Ranges |
0-10% (other ranges available) |
|
Typical Precision |
±1% of reading or ±0.04ppmv (400ppmv) whichever is greater |
|
Measurement Response Time |
0.25-2 sec
(Total system response is dependent on
fl ow rate and sample system volume) |
|
Principle of Measurement |
Tunable Diode Laser Absorption Spectroscopy
|
|
Environmental Temperature Range |
-20º to 50ºC (-4º to 122ºF)
-15º to 60º C (-5º to 140º F) optional
|
|
Sample Inlet Pressure |
10 to 25 PSIA, 10 PSIG Maximum
(70-170kPa Abs, 70kPaG maximum) |
|
Sample Cell Temperature Range |
-20º to 50ºC (-4º to 122ºF)
-15º to 60ºC (-5º to 140ºF)
|
|
Maximum Cell Pressure |
70 kPag (10 PSIg) |
| Sample Flow Rate |
0.1-10 L/min (60.2 to 20 scfh) |
| Recommended Validation |
Binary Cal Gas Bottle with Methane Background |
Electrical |
|
|
Voltage |
100-240 VAC, 50-60 Hz |
|
Max Current |
1 amp maximum @ 120 VAC
1.6A @ 24VDC, 3.2A @ 12 VDC
|
|
Communication |
Generic or Modbus RS232C
4-20mA loop
|
| Digital Outputs |
2, General Fault and Concentration Alarm |
| LCD Display |
Concentration, Cell Pressure and Temperature & Diagnostics |
Physical |
|
|
Controller Enclosure |
NEMA 4X - stainless steel standard |
|
Dimensions |
444mm H x 376mm W x 135mm H
(17.5" H x 14.8" W x 5.8" D)
|
|
Weight Approximately |
Approx 11.5Kg(25lbs) |
|
Sample Cell Dimensions |
438mm H x 108mm W
(17-1/4” H × 4-1/4” W) |
|
Sample Cell Construction |
316L Series Polished Stainless Steel standard |
| Number of Sample Cells |
1
|
| Area Classification |
CSA Certified for Class I, Division 2,
Groups ABCD, Temp Code T3C
|
Drawings
Accessories
|
02199-00001 |
Spares Kit- includes fuses, o-rings and hardware
|
|
00000-02015 |
Extended calibration range (up to 100lb/MMSCF H2O)
Other ranges available (consult factory) |
| 00000-02013 | Tropicalization - conformal coating of boards & moisture-proof sealant at electrical connections
|
|
01902-93001 |
Instruction Manuals (up to 2 copies at no charge)
|
|
01910-30017 |
Stainless Steel Tag, up to 2 lines of text
|
|
00000-02014 |
Certified As-Built Drawings - 2 sets
|
| 01902-18100 | Alarm relays, SPDT, (1) Status & (1) High Concentration not for hazardous location (non CSA)
|