4333 W. Sam Houston Pkwy N., Suite 100 Houston, TX 77043 | D: +1 713 300-2700 | TF: 800 619-2861 | F: +1 713 856-6623
Phone: 800-619-2861
E-mail: sales@spectrasensors.com

SpectraSensors SS3000e Dual Channel Gas Analyzer is capable of measuring moisture in this dual channel system which enhances cost savings by incorporating two sensors in one.
Performance |
|
|
Target Components |
Dual H2O in Natural Gas |
|
Typical Measurement Ranges - H2O |
0-422ppmv (0-20 lb/MMscf) Ranges up to 10,000ppmv available* |
|
Typical Precision - H2O |
±4ppmv |
|
Measurement Response Time |
0.25-2 seconds (Total system response dependent on flow rate and sample system volume) |
|
Principle of Measurement |
Tunable Diode Laser Absorption Spectroscopy
|
|
Environmental/Sample Temperature Range |
-20º to 50º C (-4º to 122º F)
-10º to 60º C (14º to 140º F) optional
|
|
Sample Pressure |
Maximum cell pressure 70kPaG (10 psig) |
| Sample Flow Rate |
0.5-1 L/min (1-2 scfh) |
| Validation |
H2O -Bureau of Mines Chilled Mirror or Portable TDL
|
Electrical |
|
|
Input Voltage |
100-240 VAC, 50-60 HZ |
|
Max Current (unheated) |
1 amp maximum @ 120 VAC , 1.6A @ 24VDC |
|
Max Current (heated) |
2 amps maximum @ 120 VAC |
| Communication |
Analog: Two 4-20mA Isolated, 1200 ohms @ 24 VDC max load Serial: RS232C standard, RS485 and Ethernet Optional Protocol: Modbus Gould RTU or Daniel RTU or ASCII |
| Digital Outputs |
2, General Fault and Concentration/Assignable Alarm |
| LCD Display |
Concentration, Cell Pressure, Temperature, Alarms & Diagnostics |
Physical |
|
|
Enclosure Type |
NEMA 4X Stainless Steel Enclosures |
|
Dimensions |
973, 406, 224mm (38.3, 16, 8.8 inches) |
|
Approximate Weight |
45kg (100lbs) |
|
Sample Cell Dimensions |
438 mm H x 108 mm W (17 1/4”H x 4 1/4”W)
|
|
Sample Cell Construction |
316L Series Polished Stainless Steel standard |
| Number of Sample Cells |
2
|
| Area Classification |
CSA Class 1, Division 2, Groups BCD, Temp Code T3C (T3 with Heaters) |
|
Part Number |
Description |
|
01902-17106 |
External Serial Output Cable1 |
|
01902-13000 |
Temperature Control Board |
|
01902-14000 |
4-20 mA Current Loop Board |
|
80000-02480 |
Power Supply & Filter 120/240 VAC 50/60 Hz |
|
80000-02481 |
Power Supply & Filter 12 VDC |
|
80000-02482 |
Power Supply & Filter 24 VDC |
|
31000-02151 |
Filter, Power Line 120/240 VAC, 5 amps |
|
80000-02484 |
Ethernet Converter 2-Channel |
|
80000-02486 |
RS-485 Converter 2-Channel |
|
01902-30011 |
Keypad Assembly |
|
24000-02087 |
Display Assembly |
|
45000-02002 |
Relay, DC12V SPDT 3 A/120 VAC |
|
80000-02474 |
Pressure Transducer Assembly |
|
24000-02088 |
Cable Gland/Seal, 9-Position, Roxtec RG00063090046 |
|
02199-00005 |
Kit, O-rings Viton, 2-Pass Cell1 |
|
02199-00011 |
Kit, Fuse, AC, DC1 |
|
02199-00007 |
Kit, Cleaning Tools, Optical Cell |
|
01910-30017 |
Stainless Steel Tag (up to 2 lines of text) |
|
00000-02015 |
Extended Calibration 100lbs moisture |
|
49000-02068 |
Instruction Manuals (up to 2 copies at no charge), additional copies |
61303-042S4 Ball Valve, 1/4” TF SS Swagelok SS-42GS4 61015-20074 Filter, Tee-Type, 7µm, 1/4” TF SS Swagelok SS-4TF-7 61000-02186 Filter Element, 7µm Swagelok SS-4F-K4-71 61016-71208 Membrane Separator, 1/4" NPTF SS A+ Corp 120-005-SS 61016-120X5 Membrane Replacement Filter A+ Corp 120-5X51 61341-00274 Flow Meter w/Valve 2 SLPM, 1/4" NPTF SS King 74C123G081123810 28000-02041 Rebuild Kit, Flowmeter, Viton, King 74301 61321-25254 Pressure Regulator, 25psig, 1/4” NPTF SS GO PR1-1A11A3D111 61000-02004 Pressure Gauge, 0–30 psig McDaniel SBL / Parker 9118128 62000-00006 Pressure Gauge, 0–60 psig McDaniel SCL / Parker 9118101 55032-340B6 Temperature Gage, 3”, 200F, 4” Stem, 1/2" NPT Reotemp AA-040-1-043 28011-19112 Watertight Conduit Hub 3/4 inch, Appleton HUB-75D 53000-02017 Heater 100W, 120 VAC Intertec CP Multitherm CNA 100 T3 120V 53000-02018 Heater 100W, 230 VAC Intertec CP Multitherm CNA 100 T3 230V
P53
|
|||||||||||||||||||||
HEATED SAMPLE CONDITIONING |
|||||||||||||||||||||
|
The P53 Heated Sample Conditioning System ensures performance of gas analyzer by delivering a heated sample through patented technology to the analyzer. This avoids costly errors resulting from hydrocarbon dew point drop out by the maintaining at least 30ºF above the expected hydrocarbon dew point. |
The gas sample is heated inside of the regulator before and after the pressure is reduced eliminating hydrocarbon liquid condensation caused by the Joule Thomson effect during the pressure reduction. |
||||||||||||||||||||
![]() |
|||||||||||||||||||||
P53 |
|||||||||||||||||||||
HEATED SAMPLE CONDITIONING SYSTEM |
|||||||||||||||||||||
![]() |
|||||||||||||||||||||
|
|||||||||||||||||||||
PONY |
|||||||||||||||||||
HEATED PROBE ENCLOSURE |
|||||||||||||||||||
|
The Pony® Heated Probe Enclosure is designed to provide and maintain heat for the sample probe directly at the sample point with a P53 Heated Sample Conditioning System. The Pony® enclosure is an integral part of the P53, developed to install between 6” center-to-center thread-o-lets. |
Consisting of a molded, insulated, and weatherproof clam shell design, the Pony® enclosure provides easy maintenance with steel latches and chain to prevent loss of lid. Construction selections include hot press glass fiber reinforced polyester or stainless steel. |
||||||||||||||||||
![]() |
|||||||||||||||||||
PONY
|
|||||||||||||||||||
HEATED PROBE ENCLOSURE |
|||||||||||||||||||
![]() |
|||||||||||||||||||
|
|||||||||||||||||||
To request a quote online, please enter the following and we'll get back to you promptly: