SS3000 - Dual H2O / H2O Analyzer Solution

 

 
For a dual H2O measurement, the SS3000 Single Channel Analyzer is capable of measuring two dual moisture channels using tunable diode laser technology. The two components are measured in a single analyzer. The SS3000 delivers precise, reliable measurements and dramatically reduces maintenance costs.

 

Product Code: 10805

Drawings
 

 

Application Note

 
Rapid Response Time

The SS3000 analyzer takes four measurements per second with a laser and detector and immediately averages the results. Because there is no contact with the gas, real-time measurements are not hampered by wet-up or dry-down times as with surfaced-based sensors.


Trustworthy Measurements
Dependable data is an essential element in the quest for improved safety and quality. The SS3000 analyzer delivers precise, reliable measurements using patented Tunable Diode Laser (TDL) technology developed by NASA.
 
The TDL sensor never comes into contact with the sample gas stream. The result is a sensor which does not suffer from contamination or drift due to vapor impurities such as glycol, methanol or amines.
 
The SS3000 dramatically reduces intangible but real costs associated with unreliable gas measurements. By eliminating added processing steps, detecting poor gas quality and the possibility of costly damage to equipment that can result from sensors that produce incorrect data.
 
State of the Art Technology
The analyzer works by shining a laser beam through the sample cell. The laser beam is selected to interact only with the measured compound, creating an absorption signal. The higher the concentration of H2O, the greater absorption of light and the stronger the corresponding absorption signal. Spectrum Software analyzes these absorption peaks to produce very accurate and repeatable measurements. Since the calculation is a direct, fundamental measurement, the amount of H2O present can be measured in real-time.
 
Low Cost of Ownership
Operating costs are dramatically reduced by eliminating the cost of consumables, extra sensor heads, labor and overhead associated with excessive maintenance.
 
 

« Previous | Next »

Data Sheet

Performance

Target Components

H2O/H2O in Natural Gas

Typical Measurement Ranges

0-100ppmv
0-422ppmv
0-1000ppmv (ranges up to 10,000ppmv available)

Typical Precision

±1% of reading or ±4ppmv

Measurement Response Time

Total system response is dependent on flow rate and sample system volume

Principle of Measurement

Tunable Diode Laser Absorption Spectroscopy

Environmental Temperature Range

-20º to 50ºC (-4º to 122ºF)
-15º to 60ºC (-5º to 140ºF) optional

Sample Inlet Pressure

10 PSIA, 10 PSIG Maximum

(70-170kPa Abs, 70kPaG maximum)

Sample Cell Temperature Range

-20º to 50ºC (-4º to 122ºF)
-15º to 60ºC (-5º to 140ºF)

Maximum Cell Pressure

70 kPag (10 PSIg)

Sample Flow Rate
0.1-10 L/min (60.2 to 20 scfh)
Validation
Bureau of Mines Chilled Mirror or Portable TDL
 

Electrical

Power

100-240 VAC, 50-60 Hz standard

Max Current

1 amp maximum @ 120 VAC
1.6A @ 24VDC, 3.2A @ 12 VDC

Communication

Generic or Modbus RS232C
4-20mA loop
Digital Outputs
4, General Fault and Concentration Alarms for Each Measurement Channel
LCD Display
Concentration, Cell Pressure and Temperature & Diagnostics
 

Physical

Controller Enclosure

NEMA 4X - stainless steel standard

Dimensions

444mm H x 376mm W x 135mm H
(17.5" H x 14.8" W x 5.8" D)

Weight Approximately

Approx 11.5Kg (25lbs)

Sample Cell Dimensions

438mm H x 108mm W
(17-1/4” H × 4-1/4” W)

Sample Cell Construction

316L Series Polished Stainless Steel standard

Number of Sample Cells
2 (dual channel SS3000)
Area Classification
CSA Certified for Class I, Division 2,
Groups ABCD, Temp Code T3C


Drawings

 
SS3000 Dual Channel with Stainless Steel Enclosure (NEMA 7,4X) – 2 stream
 
SS3000 with Exd aluminum enclosure (NEMA 7,4)
 
 
 

« Previous | Next »

Typical Accessories

02199-00001

Spares Kit- includes fuses, o-rings and hardware
 

00000-02015

Extended calibration range (up to 100lb/MMSCF H2O)
Other ranges available (consult factory)
 
00000-02013 Tropicalization - conformal coating of boards & moisture-proof sealant at electrical connections
 

01902-93001

Instruction Manuals (up to 2 copies at no charge)
 

01910-30017

Stainless Steel Tag, up to 2 lines of text
 

00000-02014

Certified As-Built Drawings - 2 sets
 
01902-18100 Alarm relays, SPDT, (1) Status & (1) High Concentration not for hazardous location (non CSA)
 
 
 
 

Photos

 