Skip to Content

4333 W. Sam Houston Pkwy N., Suite 100 Houston, TX 77043 | D: +1 713 300-2700 | TF: 800 619-2861 | F: +1 713 856-6623

SS2100 - H2S Analyzer Solution

 
For H2S the SS2100 Analyzer is capable of measuring  hydrogen sulfide using tunable diode laser technology. The SS2100 delivers precise, reliable measurements and dramatically reduces maintenance costs.
 
This application measures H2S with typical measurements of 0-20 ppm with a repeatability of  ± 500 ppb.

 

 Application Code: 10902

 

Drawings
 
 

Application Note

SpectraSensors SS2100 H2S Measurement Analyzer measures hydrogen sulfide in natural gas quickly and reliably without light source aging, tape replacement or ongoing recalibration.
 
Minimal maintenance
The SS2100 H2S is perfect for remote or hazardous installations because its patented, laser based technology and requires very little annual maintenance. Factory set calibration last the full lifetime of the analyzer.
 
Reliable
SpectraSensors analyzers are the first to offer truly dependable technology and simple operation for applications such as natural gas pipelines monitoring.

By its nature, the TDL (Tunable Diode Laser) detection method is not susceptible to aging effects. Excellent measurement repeatability between instruments compared to lab instruments makes the SS2100 H2S analyzer the ideal measurement tool for even the most punishing of environments.

Sensitive enough to meet the needs of critical tariff applications, the analyzer has operating ranges of 0-10 ppmv to percent levels and repeatability as low as ±500 ppbv.
 
Easy Operation
The SS2100 H2S is remarkably easy to install; connect the power, the data link and the measured gas line and the analyzer begins working. Once set up, measurements begin within minutes.
 
Versatility
The SS2100 H2S analyzer is available in many configurations. An optional measurement cell can be added to the analyzer for dual measurement H2S specifications, see the SS2000 product sheets.

 

« Previous | Next »

Data Sheet

Performance

Target Components

H2S in Natural Gas

Typical Measurement Ranges

0-20 ppm*

Typical Precision

±500ppb

Measurement Response Time

16 seconds**

Principle of Measurement

Tunable Diode Laser Absorption Spectroscopy

Environmental Temperature Range

-20º to 50ºC (-4º to 122ºF)
-15º to 60º C (-5º to 140º F) optional

Sample Inlet Pressure

130 - 340 kPaG

(20-50 PSIG) to panel

Sample Cell Temperature Range

-20º to 50ºC (-4º to 122ºF)
-15º to 60ºC (-5º to 140ºF)

Maximum Cell Pressure

70 kPag (10 PSIg)

Sample Flow Rate
3 L/min (5.4 scfh)
Recommended Validation Binary Cal Gas Bottle with Methane Background
 

Electrical

Voltage

100-240 VAC, 50-60 Hz

Max Current

2 amp maximum @ 120 VAC
1 amp maximum @ 120 VAC

Communication

Generic or Modbus RS232C
4-20mA loop
Digital Outputs
2, General Fault and Concentration Alarm
LCD Display
Concentration, Cell Pressure and Temperature
& Diagnostics
 

Physical

Controller Enclosure

NEMA 4X - stainless steel standard

Dimensions

610mm H x 1372mm W x 330mm H
(24" H x 54" W x 13" D)

Weight Approximately

Approx 91Kg (200lbs)

Sample Cell Dimensions

438mm H x 108mm W
(17-1/4” H × 4-1/4” W)

Sample Cell Construction

316L Series Polished Stainless Steel standard

Number of Sample Cells
1 (2-cell or 3-cell systems available)
Area Classification
CSA Certified for Class I, Division 2,
Groups ABCD, Temp Code T3C
* Typical Shown; custom configurations available
** Application specific; consult factory

Drawings

 

 
 
 
 

« Previous | Next »

Typical Accessories


 
 
 

Photos

 

« Previous | Next »

« Previous | Next »