Ethylene is one of the most important feedstocks for the petrochemical industry. Since it is used in the manufacture of a wide range of compounds, Ethylene has very stringent purity specifications. One of the more important steps in purifying the ethylene is removing the acetylene. There are 4 common methods of Acetylene removal in Ethylene process that have evolved over the years:
- Solvent Extraction using Dimethyl Formamide (DMF)
- Cracked Gas Train or Raw Gas Catalytic Hydrogenation
- Back End Catalytic Hydrogenation
- Front End Catalytic Hydrogenation
Application Note
Acetylene Converter UnitThe Acetylene Converter Unit consists of a series of reactors (Fig. 1) or a single vessel with multiple beds of catalyst (Fig. 2) that reacts the Acetylene with Hydrogen to form Ethylene. The reactor or bed will reduce the amount of Acetylene from several thousand ppm down to several hundred ppm. The second reactor or bed reduces the Acetylene further to the low ppm and even ppb levels. Back End Acetylene Converters are typically located downstream of the De-Ethanizer on a mixture of Ethane and Ethylene to which Hydrogen and sometimes CO are added to reduce the Acetylene.
Critical Control of Acetylene
It is critical that the Acetylene be brought down to the proper levels as it exits the final reactor; otherwise, any Acetylene present will follow the Ethylene all the way to the final product stream. But if the unit operates too aggressively, it could hydrogenate some of the Ethylene into Ethane. Measurements are made at the outlet of the Acetylene Converter Unit for feedback control strategies, and also to ensure the final product specification for Acetylene is met.
Traditional Measurement Solutions
On-line gas chromatographs have been the traditional method for monitoring the levels of Acetylene in the process streams. Unfortunately, even with the latest in chromatography techniques, the analysis can take 3-6 minutes between measurement updates. Due to the very short residence times in the reactors, this delay in measurements can easily lead to excursions in concentrations before it is detected. Also, GC’s consume carrier and flame fuel cylinder gases, as well as requiring hydrocarbon-free air for flame ionization detectors, so the consumable costs of GC’s are high.
SpectraSensors’ Solution
The SpectraSensors SS2100 is the ideal solution for this challenging application. The use of Tunable Diode Laser technology means that analysis results can be updated every second if desired. Furthermore, the high resolution that is inherent to TDL technology eliminates errors due to interferences that have hampered other spectrometric approaches. There are no consumables required for Acetylene analysis, so cost of ownership is low.
|
Component |
Unit |
Typical Concentration |
Min. for Application |
Max. for Application |
|
Acetylene (C2H2) |
ppmv |
0 - 5 |
0 |
10 |
|
Ethylene (C2H4) |
Mole% |
65 |
60 |
90 |
|
Ethane (C2H6) |
Mole% |
33 |
0 |
40 |
|
Hydrogen (H2) |
ppmv |
10 - 50 |
0
|
1000
|
|
Carbon Monoxide (CO) |
ppmv |
0.5 |
0
|
1000 |
|
Carbon Dioxide (CO2) |
ppmv |
< 1.0 |
0 |
1000 |
|
Methane (CH4) |
ppmv |
50 - 100 |
0 |
1000 |
|
Propylene (C2H6) |
ppmv |
3000 |
0 |
5000 |
|
Total |
Mole% |
100 |
|
|
The background stream composition must be specified for proper calibration and measurement performance. Specify the Normal composition, along with the minimum and maximum expected values for each component, especially water, the measured component. Other stream compositions may be allowable with approval from SpectraSensors.
Data Sheet
Performance |
|
|
Target Components |
Acetylene in Back End Acetylene Converter Inlet |
|
Typical Measurement Ranges |
0 – 5 ppm* |
|
Typical Precision |
±0.1ppm |
|
Measurement Response Time |
1 to ~60 sec |
|
Principle of Measurement |
Tunable Diode Laser Absorption Spectroscopy
Non-differential |
|
Environmental Temperature Range |
-20º to 50ºC (-4º to 122ºF)
-10º to 60ºC (14º to 140ºF) optional
|
|
Sample Inlet Pressure |
70 kPag (10 psig) typical
210 kPag (30 psig) maximum
|
|
Cell Temperature |
Maintain at 50ºC +/- 0.2ºC |
|
Maximum Cell Pressure |
70 kPag (10 PSIg) |
| Sample Flow Rate |
3 - 4 L/min (6.4 to 8.5 scfh)* |
Electrical |
|
|
Power |
100-240 VAC, 50-60 Hz standard |
|
Max Current |
Controller: 1 A @ 120 VAC |
|
Controller to Cell Cable Length |
1m standard (3m, 5m & 10m available optionally) |
|
Communication |
Current Loop Output 4-20 mA Isolated, 1200 ohms @ 24 VDC max load
Serial: ASCII Text RS232C standard
Modbus RS232C
|
| Digital Outputs |
4 ea. 12 VDC for valve operations
Scrubber (if required), Process/Val, Val 1, Val 2
5 SPDT (Form C) Dry Contact Alarms
Common Fault, Val 1 Active, Val 2 Active, Val Fail,
one user-assignable to any alarm
|
| LCD Display |
Concentration, Cell Pressure and Temperature, Diagnostic Data |
Physical |
|
|
Controller Enclosure |
NEMA 4X - 304 Stainless Steel standard |
|
Controller Dimensions |
343mm H x 305mm W x 165mm H
(13.5" H x 12" W x 6-7/16" D)*
|
|
Weight Approximately |
13.1 Kg (28.6 lbs)* |
|
Sample Cell Dimensions |
28m cell for ranges
559 mm H × 127 mm W (22” H × 5” W) |
|
Sample Cell Construction |
316L Series Polished Stainless Steel standard
SilcoNert® coated |
| Number of Sample Cells |
1 (single channel SS2100) or
2 (dual channel SS2100)
|
| Area Classification |
CSA Certified for Class I, Division 2,
Groups ABCD Temp Code T3C
|
| Dimensions with Sample System | 1678 mm H x 613 mm W x 427 mm D (66” H x 24-1/8” W x 16-13/16” D) |
| Weight with Sample System (approx.) | 68 kg (150 lbs.) |
|
* Application specific; consult factory
|
|
Typical Accessories
|
SS2100 ARM9 Electronics
|
|
|
80000-02199 |
Temperature Control Board |
|
23000-00030 |
4-20 mA Current Loop Board
|
|
80000-02203 |
Power Supply Assembly, 120/240 VAC 50/60 Hz
|
|
80000-02300 |
Power Supply Assembly, 24 VDC
|
| 01902-17106 | External Serial Output Cable |
| 01902-30011 | Keypad Assembly |
| 24601-00002 |
Display Assembly
|
| 02203-13100 |
Daughter Board
|
| 45000-02002 |
Relay, DC12V SPDT 3 A/120 VAC
|
| Herriott Cell (8m & 28m) |
|
| 02101-17102 |
Pressure Transducer Assembly
|
| 02101-17103 |
Thermistor Assembly
|
| 02199-00006 |
Spares Kit (o-rings, fuses), AC, Viton
|
| 02199-00007 |
Cleaning Kit (domestic only)
|