Application Note
Acetylene Control and ReductionIt is not practical to remove Acetylene from Ethylene by distillation. Acetylene reduction is done in an Acetylene Converter Unit. Acetylene Converter Units consists of a series of reactors or a single vessel with multiple beds of catalyst that reacts the Acetylene with hHdrogen to form Ethylene. Reduction of low ppm Acetylene concentrations inevitably reacts some of the Ethylene to Ethane, resulting in a loss of product, so it is critical to keep Acetylene contamination out of Ethylene once the initial Acetylene reduction is complete.
Traditional Measurement Solutions
On-line gas chromatographs have been the traditional method for monitoring the levels of Acetylene in Polymer Grade Ethylene. Unfortunately, even with the latest in chromatography techniques, the analysis can take 2 - 4 minutes between measurement updates. Due to the very short residence times in Ethylene feed storage vessels and the desired to measure Ethylene in flowing pipelines, this delay in measurements can easily lead to excursions in concentrations before it is detected.
SpectraSensors’ Solution
The SpectraSensors SS2100 is the ideal solution for Ethylene purity applications. The use of Tunable Diode Laser technology means that analysis results can be updated every second. Furthermore, the high resolution that is inherent to the laser eliminates errors due to interferences that have hampered other spectrometric approaches. Also, the non-contact nature of the measurement with no moving parts in the analyzer means that the analyzer is simple to operate, has no consumable parts and no routine maintenance is needed. See Fig. 3 for typical calibration performance.
In addition to Acetylene, water, another important purity measurement can be made using TDL analysis.
|
Component |
Unit |
Typical Concentration |
Min. for Application |
Max. for Applicaiton |
|
Ethylene (C2H4) |
Mole% |
99.95 |
99.9 |
100 |
|
Acetylene (C2H2) |
ppmw |
<1 |
0 |
5 |
|
Water |
ppmw |
<1 |
0 |
5 |
|
Carbon Monoxide (CO) |
ppmw |
0.5 |
0
|
3 |
|
Carbon Dioxide (CO2) |
ppmw |
< 1.0 |
0 |
5 |
|
Hydrogen (H2) |
ppmw |
<1 |
0
|
5
|
|
“Light Inerts” (C1+C2+N2) |
ppmv |
100 - 200 |
0 |
1000 |
|
Propylene (C2H6)+ |
ppmw |
3000 |
0 |
10 |
|
Total |
Mole% |
100 |
|
|
Data Sheet
Performance |
|
|
Target Components |
Acetylene in Pure Ethylene (Product or Feed) |
|
Typical Measurement Ranges |
0 – 5 ppm* |
|
Typical Precision |
± 0.1 ppm |
|
Measurement Response Time |
1 to ~60 sec* |
|
Principle of Measurement |
Tunable Diode Laser Absorption Spectroscopy
Non-differential |
|
Environmental Temperature Range |
-20º to 50ºC (-4º to 122ºF)
-10º to 60ºC (14º to 140ºF) optional
|
|
Sample Inlet Pressure |
70 kPag (10 psig) typical
210 kPag (30 psig) maximum
|
|
Cell Temperature |
Maintain at 50ºC +/- 0.2ºC |
|
Maximum Cell Pressure |
70 kPag (10 PSIg) |
|
Sample Flow Rate
|
3 - 4 L/min (6.4 to 8.5 scfh)* |
| Validation |
Certified blend of C2H2 in Nitrogen balance |
Electrical |
|
|
Power |
100-240 VAC, 50-60 Hz standard |
|
Max Current |
Controller: 1 A @ 120 VAC |
|
Controller to Cell Cable Length |
1m standard (3m, 5m & 10m available optionally) |
|
Communication |
Current Loop Output 4-20 mA Isolated,
1200 ohms @ 24 VDC max load
Serial: ASCII Text RS232C standard
Modbus RS232C
|
| Digital Outputs |
4 ea. 12 VDC for valve operations
Scrubber (if required), Process/Val, Val 1, Val 2
5 SPDT (Form C) Dry Contact Alarms
Common Fault, Val 1 Active, Val 2 Active, Val Fail,
one user-assignable to any alarm
|
| LCD Display |
Concentration, Cell Pressure and Temperature, Diagnostic Data |
Physical |
|
|
Controller Enclosure |
NEMA 4X - 304 Stainless Steel standard |
|
Controller Dimensions |
343mm H x 305mm W x 165mm H
(13.5" H x 12" W x 6-7/16" D)*
|
|
Weight Approximately |
13.1 Kg (28.6 lbs)* |
|
Sample Cell Dimensions |
28m cell
559 mm H × 127 mm W (22” H × 5” W) |
|
Sample Cell Construction |
316L Series Polished Stainless Steel standard
SilcoNert® coated |
| Number of Sample Cells |
1 (single channel SS2100) or
2 (dual channel SS2100)
|
| Area Classification |
CSA Certified for Class I, Division 2,
Groups ABCD Temp Code T3C
|
| Dimensions with Sample System | 1678 mm H x 613 mm W x 427 mm D (66” H x 24-1/8” W x 16-13/16” D) |
| Weight with Sample System (approx.) | 68 kg (150 lbs.) |
|
* Application specific; consult factory
|
|
Typical Accessories
|
SS2100 ARM9 Electronics
|
|
|
80000-02199 |
Temperature Control Board |
|
23000-00030 |
4-20 mA Current Loop Board
|
|
80000-02203 |
Power Supply Assembly, 120/240 VAC 50/60 Hz
|
|
80000-02300 |
Power Supply Assembly, 24 VDC
|
| 01902-17106 | External Serial Output Cable |
| 01902-30011 | Keypad Assembly |
| 24601-00002 |
Display Assembly
|
| 02203-13100 |
Daughter Board
|
| 45000-02002 |
Relay, DC12V SPDT 3 A/120 VAC
|
| Herriott Cell (8m & 28m) |
|
| 02101-17102 |
Pressure Transducer Assembly
|
| 02101-17103 |
Thermistor Assembly
|
| 02199-00006 |
Spares Kit (o-rings, fuses), AC, Viton
|
| 02199-00007 |
Cleaning Kit (domestic only)
|