SS2100 - C2H2 Ethylene Purity (Product or Polymer Feed) Analyzer Solution

 
Ethylene is one of the most important feedstocks for the petrochemical industry. Since it is used in the manufacture of a wide range of compounds, Ethylene has very stringent purity specifications especially for Polymer Grade Ethylene as Acetylene poisons polyethylene catalysts.  Sample points for Acetylene in Polymer Grade Ethylene are either in the Ethylene plant product or in a Polyethylene plant feed.  Large quantities of Ethylene are transported via pipelines and stored in underground salt caverns.  In both cases the purity of the Ethylene must be verified at the point of custody transfer.

 

Product Code: 54911
 
 
 
 
 

Application Note

 Acetylene Control and Reduction
It is not practical to remove Acetylene from Ethylene by distillation. Acetylene reduction is done in an Acetylene Converter Unit. Acetylene Converter Units consists of a series of reactors or a single vessel with multiple beds of catalyst that reacts the Acetylene with hHdrogen to form Ethylene. Reduction of low ppm Acetylene concentrations inevitably reacts some of the Ethylene to Ethane, resulting in a loss of product, so it is critical to keep Acetylene contamination out of Ethylene once the initial Acetylene reduction is complete.

 Traditional Measurement Solutions
On-line gas chromatographs have been the traditional method for monitoring the levels of Acetylene in Polymer Grade Ethylene. Unfortunately, even with the latest in chromatography techniques, the analysis can take 2 - 4 minutes between measurement updates. Due to the very short residence times in Ethylene feed storage vessels and the desired to measure Ethylene in flowing pipelines, this delay in measurements can easily lead to excursions in concentrations before it is detected.

 SpectraSensors’ Solution
The SpectraSensors SS2100 is the ideal solution for Ethylene purity applications. The use of Tunable Diode Laser technology means that analysis results can be updated every second. Furthermore, the high resolution that is inherent to the laser eliminates errors due to interferences that have hampered other spectrometric approaches. Also, the non-contact nature of the measurement with no moving parts in the analyzer means that the analyzer is simple to operate, has no consumable parts and no routine maintenance is needed. See Fig. 3 for typical calibration performance.

In addition to Acetylene, water, another important purity measurement can be made using TDL analysis.
 
Typical Stream Composition

Component

Unit

Typical Concentration

Min. for Application

Max. for Applicaiton

Ethylene (C2H4)

Mole%

99.95

99.9

100

Acetylene (C2H2)

ppmw

<1

0

5

Water

ppmw

<1

0

5

Carbon Monoxide (CO)

ppmw

0.5

0

 

3

Carbon Dioxide (CO2)

ppmw

< 1.0

0

5

Hydrogen (H2)

ppmw

<1

0

 

5

 

“Light Inerts”

(C1+C2+N2)

ppmv

100 - 200

0

1000

Propylene (C2H6)+

ppmw

3000

0

10

Total

Mole%

100

 

 

 

« Previous | Next »

Data Sheet


Performance

Target Components

Acetylene in Pure Ethylene (Product or Feed)

Typical Measurement Ranges

0 – 5 ppm*

Typical Precision

 ± 0.1 ppm

Measurement Response Time

1 to ~60 sec*

Principle of Measurement

Tunable Diode Laser Absorption Spectroscopy
Non-differential

Environmental Temperature Range

-20º to 50ºC (-4º to 122ºF)
-10º to 60ºC (14º to 140ºF) optional

Sample Inlet Pressure

70 kPag (10 psig) typical

210 kPag (30 psig) maximum


Cell Temperature

Maintain at 50ºC +/- 0.2ºC

Maximum Cell Pressure

70 kPag (10 PSIg)

Sample Flow Rate
 
3 - 4 L/min (6.4 to 8.5 scfh)*
Validation
Certified blend of C2H2 in Nitrogen balance
 

Electrical

Power

100-240 VAC, 50-60 Hz standard

Max Current

Controller: 1 A @ 120 VAC

Controller to Cell Cable Length

1m standard (3m, 5m & 10m available optionally)

Communication

Current Loop Output 4-20 mA Isolated,
1200 ohms @ 24 VDC max load
Serial: ASCII Text RS232C standard
Modbus RS232C
Digital Outputs
4 ea. 12 VDC for valve operations
Scrubber (if required), Process/Val, Val 1, Val 2
5 SPDT (Form C) Dry Contact Alarms
Common Fault, Val 1 Active, Val 2 Active, Val Fail,
one user-assignable to any alarm 
LCD Display
Concentration, Cell Pressure and Temperature, Diagnostic Data
 

Physical

Controller Enclosure

NEMA 4X - 304 Stainless Steel standard

Controller Dimensions

343mm H x 305mm W x 165mm H
(13.5" H x 12" W x 6-7/16" D)*

Weight Approximately

13.1 Kg (28.6 lbs)*

Sample Cell Dimensions

28m cell
559 mm H × 127 mm W
(22” H × 5” W)

Sample Cell Construction

316L Series Polished Stainless Steel standard
SilcoNert® coated
Number of Sample Cells
1 (single channel SS2100) or
2 (dual channel SS2100)
Area Classification
CSA Certified for Class I, Division 2,
Groups ABCD Temp Code T3C
 ll 2G Ex d llB+H2 T5; Tamb : -20 ÷ +60 °C
Dimensions with Sample System 1678 mm H x 613 mm W x 427 mm D
(66” H x 24-1/8” W x 16-13/16” D)
Weight with Sample System (approx.) 68 kg (150 lbs.)

* Application specific; consult factory

« Previous | Next »

Drawings

Typical Accessories


SS2100 ARM9 Electronics
 

80000-02199

Temperature Control Board

23000-00030

4-20 mA Current Loop Board
 

80000-02203

Power Supply Assembly, 120/240 VAC 50/60 Hz
 

80000-02300

Power Supply Assembly, 24 VDC
 

 

01902-17106 External Serial Output Cable
01902-30011 Keypad Assembly
24601-00002
Display Assembly
 
02203-13100
Daughter Board
 
45000-02002
Relay, DC12V SPDT 3 A/120 VAC
 

Herriott Cell (8m & 28m)
02101-17102
Pressure Transducer Assembly
 
02101-17103
Thermistor Assembly
 
02199-00006
Spares Kit (o-rings, fuses), AC, Viton
 
02199-00007
Cleaning Kit (domestic only)
 
 