Ethylene is one of the most important feedstocks for the petrochemical industry. Since it is used in the manufacture of a wide range of compounds, Ethylene has very stringent purity specifications especially for Polymer Grade Ethylene as water poisons polyethylene catalysts. Sample points for water in Polymer Grade Ethylene are either in the Ethylene plant product or in a Polyethylene plant feed. Large quantities of Ethylene are transported via pipelines and stored in underground salt caverns. In both cases, the purity of the Ethylene must be verified at the point of custody transfer.
Application Note
Water Control and ReductionAs Ethylene is separated from Ethane in a cryogenic distillation, the cracked gas must be dried, typically over Molecular Sieves and the Ethylene product is quite dry. However, from that point on, Ethylene can pick of traces of water from transportation in pipelines and ships, storage in salt caverns, and other handling. Molecular Sieve dryers can be used to dry off-spec Ethylene.
Traditional Measurement Solutions
Vibrating crystal, AlO3 capacitance probes and electrolytic moisture analyzers have been the traditional methods for monitoring the levels of water in Polymer Grade Ethylene. These methods suffer from slow wet-up and dry-down time, slow drift in calibration and sensitivity to contamination Due to the very short residence times in Ethylene feed storage vessels and the desired to measure Ethylene in flowing pipelines, the delay in measurements can easily lead to excursions in concentrations before it is detected.
SpectraSensors' Solution
The SpectraSensors SS2100 is the ideal solution for Ethylene purity applications. The use of Tunable Diode Laser technology means that analysis results can be updated every second. Furthermore, the high resolution that is inherent to the laser eliminates errors due to interferences which have hampered other spectrometric approaches. Also, the non-contact nature of the measurement with no moving parts in the analyzer means that the analyzer is simple to operate and little routine maintenance is needed.
Typical Stream Composition
|
Component |
Unit |
Typical Concentration |
Min. for Application |
Max. for Applicaiton |
|
Ethylene (C2H4) |
Mole% |
99.95 |
99.9 |
100 |
|
Water |
ppmw |
<1 |
0 |
5 |
|
Acetylene (C2H2) |
ppmw |
<1 |
0 |
5 |
|
Carbon Monoxide (CO) |
ppmw |
0.5 |
0
|
3 |
|
Carbon Dioxide (CO2) |
ppmw |
< 1.0 |
0 |
5 |
|
Hydrogen (H2) |
ppmw |
<1 |
0
|
5
|
|
“Light Inerts” (C1+C2+N2) |
ppmv |
100 - 200 |
0 |
1000 |
|
Propylene (C2H6)+ |
ppmw |
3000 |
0 |
10 |
|
Total |
Mole% |
100 |
|
|
Data Sheet
Performance |
|
|
Target Components |
Water in Pure Ethylene |
|
Typical Measurement Ranges |
0 – 10 ppm* |
|
Typical Precision |
± 0.05 ppm or 2% of Full Scale |
|
Measurement Response Time |
1 to ~60 sec* |
|
Principle of Measurement |
Differential Tunable Diode Laser Absorption Spectroscopy
(H2O Scrubber included) |
|
Environmental Temperature Range |
-20º to 50ºC (-4º to 122ºF)
-10º to 60ºC (14º to 140ºF) optional
|
|
Sample Inlet Pressure |
70 kPag (10 psig) typical
210 kPag (30 psig) maximum
|
|
Cell Temperature |
Maintain at 50ºC +/- 0.2ºC |
|
Maximum Cell Pressure |
70 kPag (10 PSIg) |
| Sample Flow Rate |
3 - 4 L/min (6.4 to 8.5 scfh)*
|
| Validation |
A certified blend of H2O in Nitrogen balance
is diluted with dried sample in the sample
conditioning system under flow control
|
Electrical |
|
|
Power |
100-240 VAC, 50-60 Hz standard |
|
Max Current |
Controller: 1 A @ 120 VAC |
|
Controller to Cell Cable Length |
1m standard (3m, 5m & 10m available optionally) |
|
Communication |
Current Loop Output 4-20 mA Isolated, 1200 ohms @ 24 VDC max load
Serial: ASCII Text RS232C standard
Modbus RS232C
|
| Digital Outputs |
4 ea. 12 VDC for valve operations
Scrubber (if required), Process/Val, Val 1, Val 2
5 SPDT (Form C) Dry Contact Alarms
Common Fault, Val 1 Active, Val 2 Active, Val Fail,
one user-assignable to any alarm
|
| LCD Display |
Concentration, Cell Pressure and Temperature, Diagnostic Data |
Physical |
|
|
Controller Enclosure |
NEMA 4X - 304 Stainless Steel standard |
|
Controller Dimensions |
343mm H x 305mm W x 165mm H
(13.5" H x 12" W x 6-7/16" D)*
|
|
Weight Approximately |
13.1 Kg (28.6 lbs)* |
|
Sample Cell Dimensions |
28m cell for ranges <100 ppm
559 mm H × 127 mm W (22” H × 5” W) |
|
Sample Cell Construction |
316L Series Polished Stainless Steel standard
Siltek® coated |
| Number of Sample Cells |
1 (single channel SS2100) or
2 (dual channel SS2100)
|
| Area Classification |
CSA Certified for Class I, Division 2,
Groups ABCD Temp Code T3C
|
| Dimensions with Sample System | 1678 mm H x 613 mm W x 427 mm D (66” H x 24-1/8” W x 16-13/16” D) |
| Weight with Sample System (approx.) | 68 kg (150 lbs.) |
|
* Application specific; consult factory
|
|
Typical Accessories
|
SS2100 ARM9 Electronics
|
|
|
80000-02199 |
Temperature Control Board |
|
23000-00030 |
4-20 mA Current Loop Board
|
|
80000-02203 |
Power Supply Assembly, 120/240 VAC 50/60 Hz
|
|
80000-02300 |
Power Supply Assembly, 24 VDC
|
| 01902-17106 | External Serial Output Cable |
| 01902-30011 | Keypad Assembly |
| 24601-00002 |
Display Assembly
|
| 02203-13100 |
Daughter Board
|
| 45000-02002 |
Relay, DC12V SPDT 3 A/120 VAC
|
| Herriott Cell (8m & 28m) |
|
| 02101-17102 |
Pressure Transducer Assembly
|
| 02101-17103 |
Thermistor Assembly
|
| 02199-00006 |
Spares Kit (o-rings, fuses), AC, Viton
|
| 02199-00007 |
Cleaning Kit (domestic only)
|