In Ethylene production one of the most important steps is removal of Hydrogen Sulfide and Carbon Dioxide from the process stream. Both H2S and CO2 are poisons to polymerization catalysts, and the allowable levels of both components are less than 5 ppmw in polymer grade Ethylene.
Application Note
Caustic Wash TowerHydrogen Sulfide is required in Ethylene cracking furnaces to passivate the coil metallurgy, and also to limit the formation of CO2 and CO. Any CO2 entering the furnaces with the feedstock, CO2 formed by the water shift reaction and CO2 formed by decoking must be removed in the Caustic Wash Tower. CO2 and H2S are removed in the Caustic Wash Tower (Fig. 1) which is typically located after the third stage of compression on the cracked gas. The acid gases are absorbed in a dilute caustic solution.
Critical Control of Carbon Dioxide
As Carbon Dioxide and H2S concentrations can change depending on the feedstock and operating conditions of the process, the loading on the Caustic Wash Tower is highly variable. Measurement of CO2 at the Caustic Wash Tower Inlet is critical to the unit caustic control.
Traditional Measurement Solutions
On-line gas chromatographs have been the traditional method for monitoring the levels of CO2 in cracked gas streams. Unfortunately, even with the latest in chromatography techniques, the analysis can take 3-6 minutes between measurement updates. Due to rapid changes in the acid gas concentration, this delay may not be acceptable. Also, gas chromatographs consume carrier and flame fuel cylinder gases, as well as requiring hydrocarbon-free air for flame ionization detectors, so the consumable costs of GC’s are high. Spectroscopic methods using broadband light sources and narrow band pass filters are prone to interferences from changing background concentrations.
SpectraSensors’ Solution
The SpectraSensors SS2100 offers a new solution to this control measurement. The use of Tunable Diode Laser technology means that analysis results can be updated every second if desired. Furthermore, the high resolution that is inherent to TDL technology eliminates errors due to interferences that have hampered other spectrometric approaches. There are no consumables required for CO2 analysis at normal concentrations, so cost of ownership is low.
|
Component |
Unit |
Typical Concentration |
Min. for Application |
Max. for Application |
|
Carbon Dioxide (CO2) |
ppmv |
200 |
10 |
500 |
|
Hydrogen Sulfide (H2S) |
Ppmv |
500 |
0 |
1000 |
|
Hydrogen (H2) |
Mole% |
25 |
15
|
30
|
|
Methane (CH4) |
Mole% |
20 |
10 |
30 |
|
Ethane (C2H6) |
Mole% |
15 |
10 |
30 |
|
Ethylene (C2H4) |
Mole% |
25 |
20 |
40 |
|
Acetylene (C2H2) |
Mole% |
0.3 |
0 |
0.5 |
|
Propylene (C2H6) |
Mole% |
7.5 |
0 |
15 |
|
Propane (C3H8) |
Mole% |
7.5 |
0 |
15 |
|
Methyl Acetylene (Propyne C3H4) |
Mole% |
0.03 |
0 |
0.1 |
|
Propadiene (C3H4) |
Mole% |
0.02 |
0 |
0.1 |
|
Carbon Monoxide (CO) |
Mole% |
0.05 |
0
|
0.1 |
|
Butanes |
Mole% |
0.05 |
0 |
0.1 |
|
Butenes |
Mole% |
0.3 |
0 |
0.5 |
|
1,3-Butadiene |
Mole% |
0.5 |
0 |
1 |
|
C5+ |
Mole% |
0.1 |
0 |
0.5 |
|
Total |
Mole% |
100 |
|
|
The background stream composition must be specified for proper calibration and measurement performance. Specify the Normal composition, along with the minimum and maximum expected values for each component, especially water, the measured component. Other stream compositions may be allowable with approval from SpectraSensors.
Data Sheet
Performance |
|
|
Target Components |
CO2 in Caustic Wash Tower Feed |
|
Typical Measurement Ranges |
0 – 500 ppm* |
|
Typical Precision |
±2% of Full Scale |
|
Measurement Response Time |
1 to ~60 sec* |
|
Principle of Measurement |
Tunable Diode Laser Absorption Spectroscopy
Non-differential |
|
Environmental Temperature Range |
-20º to 50ºC (-4º to 122ºF)
-10º to 60ºC (14º to 140ºF) optional
|
|
Sample Inlet Pressure |
70 kPag (10 psig) typical
210 kPag (30 psig) maximum
|
|
Cell Temperature |
Maintain at 50ºC +/- 0.2ºC |
|
Maximum Cell Pressure |
70 kPag (10 PSIg) |
|
Sample Flow Rate
|
3 - 4 L/min (6.4 to 8.5 scfh)* |
| Validation |
Certified blend of CO2 in Nitrogen balance |
Electrical |
|
|
Power |
100-240 VAC, 50-60 Hz standard |
|
Max Current |
Controller: 1 A @ 120 VAC |
|
Controller to Cell Cable Length |
1m standard (3m, 5m & 10m available optionally) |
|
Communication |
Current Loop Output 4-20 mA Isolated,
1200 ohms @ 24 VDC max load
Serial: ASCII Text RS232C standard
Modbus RS232C
|
| Digital Outputs |
4 ea. 12 VDC for valve operations
Scrubber (if required), Process/Val, Val 1, Val 2
5 SPDT (Form C) Dry Contact Alarms
Common Fault, Val 1 Active, Val 2 Active, Val Fail,
one user-assignable to any alarm
|
| LCD Display |
Concentration, Cell Pressure and Temperature, Diagnostic Data |
Physical |
|
|
Controller Enclosure |
NEMA 4X - 304 Stainless Steel standard |
|
Controller Dimensions |
343mm H x 305mm W x 165mm H
(13.5" H x 12" W x 6-7/16" D)*
|
|
Weight Approximately |
13.1 Kg (28.6 lbs)* |
|
Sample Cell Dimensions |
8m Herriot Cell
559 mm H × 127 mm W (22” H × 5” W) |
|
Sample Cell Construction |
316L Series Polished Stainless Steel standard
SilcoNert® coated |
| Number of Sample Cells |
1 (single channel SS2100) or
2 (dual channel SS2100)
|
| Area Classification |
CSA Certified for Class I, Division 2,
Groups ABCD Temp Code T3C
|
| Dimensions with Sample System | 1678 mm H x 613 mm W x 427 mm D (66” H x 24-1/8” W x 16-13/16” D) |
| Weight with Sample System (approx.) | 68 kg (150 lbs.) |
|
* Application specific; consult factory
|
|
Typical Accessories
|
SS2100 ARM9 Electronics
|
|
|
80000-02199 |
Temperature Control Board |
|
23000-00030 |
4-20 mA Current Loop Board
|
|
80000-02203 |
Power Supply Assembly, 120/240 VAC 50/60 Hz
|
|
80000-02300 |
Power Supply Assembly, 24 VDC
|
| 01902-17106 | External Serial Output Cable |
| 01902-30011 | Keypad Assembly |
| 24601-00002 |
Display Assembly
|
| 02203-13100 |
Daughter Board
|
| 45000-02002 |
Relay, DC12V SPDT 3 A/120 VAC
|
| Herriott Cell (8m & 28m) |
|
| 02101-17102 |
Pressure Transducer Assembly
|
| 02101-17103 |
Thermistor Assembly
|
| 02199-00006 |
Spares Kit (o-rings, fuses), AC, Viton
|
| 02199-00007 |
Cleaning Kit (domestic only)
|