Product Code: 65201
Application Note
Bulk Compressed GasesMoisture is one of the most important impurities in high purity gases such as Nitrogen, Helium Hydrogen, Argon and Carbon Dioxide. These gases are frequently used because water from the atmosphere can damage equipment or a process.
Critical Control of Moisture
Some of these bulk gases are manufactured by cryogenic processes. Others are dried using fixed bed desiccant driers as part of the process. In any case, the moisture content of the gas must be measured to ensure purity of the final product.
Traditional Measurement Solutions
Aluminum Oxide capacitance probes and vibrating quartz crystal moisture analyzers have been the traditional method for monitoring the levels of moisture at desiccant dryer outlets, or in the final product. Unfortunately, capacitance probes change their response over time even without contamination, and since they cannot be recalibrated in the field, they must be replaced. It’s not uncommon for moisture probes to be replaced multiple times a year. Quartz crystal moisture meters have slow wet-up and dry-down times that are slow to detect break-through from the dryers.
SpectraSensors’ Solution
SpectraSensors SS2100 is the ideal solution for dry bulk gas applications. Tunable Diode Laser technology means that measurement interferences from other infrared absorbing compounds are avoided. There are no wet-up or dry-down delays, resulting in fast updates even when the concentration changes dramatically (Fig. 1). The calibration is stable over the life of the analyzer. There are no consumable parts or supplies. The analyzer is virtually maintenance free and has a low cost of ownership.
Validation
|
Component |
Unit |
Typical Concentration |
Min. for Application |
Max. for Application |
|
H2O |
ppmv |
<10 |
0.05 |
100 |
|
Nitrogen (N2) |
Mole% |
99.99+ |
98.0 |
100 |
|
Helium (He) |
Mole% |
99.99+ |
98.0 |
100 |
|
Hydrogen (H2) |
Mole% |
99.99+ |
98.0 |
100 |
|
Argon (Ar) |
Mole% |
99.99+ |
98.0 |
100 |
|
Carbon Dioxide (CO2) |
Mole% |
99.99+ |
98.0 |
100 |
|
Total |
Mole% |
100 |
|
|
The background stream composition must be specified for proper calibration and measurement performance. If other than a high purity gas, please specify the normal composition, along with the minimum and maximum expected values for each component, especially water, the measured component. Other stream compositions may be allowable with approval from SpectraSensors.
Data Sheet
Performance |
|
|
Target Components |
H2O in Compressed Gases (N2/H3/H2/Air/CO2) Specify Background Gas |
|
Typical Measurement Ranges |
0 – 25ppm 0 – 10 ppm (other ranges available by request)* |
|
Typical Precision |
±0.05ppm* |
|
Measurement Response Time |
1 to ~60 sec* |
|
Principle of Measurement |
Tunable Diode Laser Absorption Spectroscopy
Non-differential |
|
Environmental Temperature Range |
-20º to 50ºC (-4º to 122ºF)
-10º to 60ºC (14º to 140ºF) optional
|
|
Sample Inlet Pressure |
70 kPag (10 psig) typical
210 kPag (30 psig) maximum
|
|
Cell Temperature |
Maintain at 50ºC +/- 0.2ºC |
|
Maximum Cell Pressure |
70 kPag (10 PSIg) |
| Sample Flow Rate |
3 - 4 L/min (6.4 to 8.5 scfh)*
|
| Validation |
A certified blend of H2O in Nitrogen balance
is diluted with dried sample in the sample
conditioning system under flow control
|
Electrical |
|
|
Power |
100-240 VAC, 50-60 Hz standard |
|
Max Current |
Controller: 1 A @ 120 VAC |
|
Controller to Cell Cable Length |
1m standard (3m, 5m & 10m available optionally) |
|
Communication |
Current Loop Output 4-20 mA Isolated,
1200 ohms @ 24 VDC max load
Serial: ASCII Text RS232C standard
Modbus RS232C
|
| Digital Outputs |
4 ea. 12 VDC for valve operations
Scrubber (if required), Process/Val, Val 1, Val 2
5 SPDT (Form C) Dry Contact Alarms
Common Fault, Val 1 Active, Val 2 Active, Val Fail,
one user-assignable to any alarm
|
| LCD Display |
Concentration, Cell Pressure and Temperature, Diagnostic Data |
Physical |
|
|
Controller Enclosure |
NEMA 4X - 304 Stainless Steel standard |
|
Controller Dimensions |
343mm H x 305mm W x 165mm H
(13.5" H x 12" W x 6-7/16" D)*
|
|
Weight Approximately |
13.1 Kg (28.6 lbs)* |
|
Sample Cell Dimensions |
28m Herriot Cell
559 mm H × 127 mm W (22” H × 5” W) |
|
Sample Cell Construction |
316L Series Polished Stainless Steel standard
SilcoNert® coated |
| Number of Sample Cells |
1 (single channel SS2100) or
2 (dual channel SS2100)
|
| Area Classification |
CSA Certified for Class I, Division 2,
Groups ABCD Temp Code T3C
|
| Dimensions with Sample System | 1678 mm H x 613 mm W x 427 mm D (66” H x 24-1/8” W x 16-13/16” D) |
| Weight with Sample System (approx.) | 68 kg (150 lbs.) |
|
* Application specific; consult factory
|
|
Typical Accessories
|
SS2100 ARM9 Electronics
|
|
|
80000-02199 |
Temperature Control Board |
|
23000-00030 |
4-20 mA Current Loop Board
|
|
80000-02203 |
Power Supply Assembly, 120/240 VAC 50/60 Hz
|
|
80000-02300 |
Power Supply Assembly, 24 VDC
|
| 01902-17106 | External Serial Output Cable |
| 01902-30011 | Keypad Assembly |
| 24601-00002 |
Display Assembly
|
| 02203-13100 |
Daughter Board
|
| 45000-02002 |
Relay, DC12V SPDT 3 A/120 VAC
|
| Herriott Cell (8m & 28m) |
|
| 02101-17102 |
Pressure Transducer Assembly
|
| 02101-17103 |
Thermistor Assembly
|
| 02199-00006 |
Spares Kit (o-rings, fuses), AC, Viton
|
| 02199-00007 |
Cleaning Kit (domestic only)
|